Share this post:
RNase Assay Kit (Fluorescence)
RNase Assay Kit (Fluorescence)

  • 542
  • 543
192T
HBP003002
48T
HBP003003

Cat. No: HBP003002、HBP003003 

Storage transportation conditions and expiry date:  

1. Transported in -25 ~ -15℃.
2. The different components of the kit are stored separately according to the temperature:

3. Store the unopened kit at for 12 months.
4. Store the kit for 6 months after opening. It is recommended to aliquot the RNA probe solution according to the single use amount to avoid light and repeated freezing and thawing


Technical Support


RNASE ASSAY KIT (FLUORESCENCE)


It is used to detect RNase contamination in the environment and materials in scientific research and production processes.

Application


It is used to detect RNase contamination in the environment and materials in scientific research and production processes.

Characteristics of Products


High sensitivity: the detection limit for RNase A is as low as 0. 313pg/mL or about 1. 56×10< sup> -9< /sup> U/μL which is only 1/8 of the best competitor kit.
High precision: intra-batch CV≤10%, inter-batch CV≤15%.
Good stability: valid for one year at -20°C. After the RNA probe solution is prepared, aliquot and store at -20°C.

Precision and Stability of Hyzmes RNase Assay Kit

Parameter 20220106 20220316 20220425
Inner-batch CV Low-value sample: 9. 4%
High-value sample: 5. 5%
Low-value sample: 6. 2%
High-value sample: 6. 2%
Low-value sample: 8. 4%
High-value sample: 8. 3%
Inter-batch CV Low-value sample: 10. 3%
High-value sample: 5. 4%
Stability The probe lyophilized powder is stored at 37°C for 7 days, the deviation from day 0 meets the requirements; After the probe lyophilized powder was dissolved, it was stored at -20°C for 182 days, the test performance meets the requirements (continuous monitoring).

 

Sensitivity higher than gel electrophoresis method


The Hzymes RNase Assay kit was in comparison with the nucleic acid hydrolysis-gel electrophoresis method, the same low-concentration enzyme samples were measured as shown in the figure below.



Hyzmes RNase Assay kit measures 2. 5×10*  U/μL RNase A,   which can be clearly distinguished from 0 concentration.






Determination of RNase A by electrophoresis.   No significant difference between 1st ~8th channels (5×10°U/μL) and 17th,   18th channels (2. 5×10°  U/L), 19th channel (negative control).


Sensitivity: Haymes RNase Assay kit VS competitor kit 


Compared with one of the best competitor kits (fluorescence), the limit of detection of Haymes RNase Assay kit is only 1/8 of that of the competitor kit.



When the sample concentration is 3. 13×10-9 U/μL (low-value sample), the signal intensity of the competitor's RNase detection kit is indistinguishable from the negative control.



When the sample concentration is 3. 13×10-9 U/μL (low-value sample), the signal intensity of the Haymes' RNase Assay Kit is clearly distinguishable from the negative control.



When the sample concentration of RNase A is from 1×10-6 to 10×10-6  pg/mL, the Signal-to-noise (S/N) values of the competitor's DNase detection kit is between 1 to 2, however the signal intensity of Haymes' DNase Assay Kit is positive correlated to the concentration of RNase I.

Composition of the Hzymes DNase Assay Kit


No. Name 192T 48T
01 10×reaction solution 2. 0mL 0. 5mL
02 RNA probe 1 tube 1 tube
03 TE buffer 2. 0mL 0. 5mL
04 RNase A standard (2U/μL) 20μL 10μL
05 Standard Dilution Buffer 12mL 6mL
06 DNase & RNase-free water 25mL 25mL
07 DNase RNase scavenger 50mL 50mL
Note: The standard is RNase A. The unit for RNase A activity is defined as the amount of enzyme that increases 1. 0 in the absorbance at 260 nm of yeast RNA at pH 5. 0 and 37 °C [3], 50 RNase A activity units are equivalent to 1 Kunitz unit [4].

Equipment and consumables required


Fluorescence microplate reader (including ex/em=535/575nm wavelength).
DNase & RNase-free pipettes and tips.
DNase & RNase-free EP tube.
DNase & RNase-free black non-transparent 96-well plate.

Preparation before experiment (check the user guide for more details)


01. Take out the Haymes RNase Assay kit and equilibrate to room temperature (18~25°C), shake and mix the components then centrifuge immediately.
02. Centrifuge the RNA probe, add TE buffer to dissolve the powder. Aliquot this RNA probe storage solution; store at-25~15°C away from light and avoid repeated freezing and thawing.
03. Set the instrument parameters: Temperature 37°C, endpoint mode, excitation wavelength/emission wavelength 535/575nm, shaking plate 10~15s before detecting.

04. Set the gain to auto-scale if possible, if the microplate reader supports kinetic detection mode, it is recommended to use the kinetic detection mode.

05. Qualitative detection: select four wells on a 96-well plate: 2μL of RNase A standard solution (10mg/mL) is diluted with standard dilution Buffer and DNase & RNase-free water to 5pg/mL (2. 5×10⁶ U/μL). Take 80μL as the positive control; 80μL of DNase & RNase-free water as the negative control, 80 μL samples were prepared in duplicate (if insufficient, dilute to 80 μL with DNase & RNase-free water). Add 10 μL RNA probe working solution and 10 μL 10×reaction solution to each well. Read the fluorescence signal value RFU0 immediately; Incubate the plate in the dark at 37°C for 30min, then read the value RFU30 immediately.

If RFU30 (sample to be tested) ≥ 2 × RFU0 (sample to be tested), it is considered that the sample to be tested is contaminated with RNase.

06. Quantitative detection: Take 2μL of RNase A standard solution (10mg/mL) and dilute it with standard dilution buffer and DNase-free water to 2. 5×10< sup> -9< /sup> mg/μL, 1. 25×10< sup> -9< /sup> mg/mL, 6. 25×10< sup> -10< /sup> mg/mL, 3. 13×10< sup> -10< /sup> mg/mL as standards; DNase & RNase-free water as the negative control; follow the same procedures of the qualitative detection, calculate ΔRFU=RFU30-RFU0, with ΔRFU as the ordinate, and the standard RNase A concentration as the abscissa (perform a linear fit, find the fitting equation y=x+b, put ΔRFU (sample to be tested) into the equation as y, identify x, multiplied by the value of sample pre-dilution, it is the approximate concentration value of the sample to be tested. )

Q& A


Q1: What is the reaction temperature and time of the RNase Assay kit?



Please incubate the reaction system at a constant temperature of 37 °C for 30 minutes.

Q2: When could false positive or false negative results or inaccurate quantitative results occur?


a. Gel buffers, high concentration of viscous substances, surfactants and dark solutions may interfere with the luminescence of fluorophores.
b. If the sample solution to be tested contains substances that inhibit DNase activity, the result of the measurement is the overall enzyme activity of the sample solution, not the activity of the enzymes within it. These substances include:
- High ionic strength solutions (eg. 5M NaCl, 20x SSC, 3M sodium acetate, etc. )
- Buffers with pH< 4 or pH> 9.
- Chaotropic agents, detergents, chelating agents or any solution that denatures proteins (eg. SDS, guanidine thiocyanate, urea, EDTA, etc. )
c. Solutions that cause chemical instability of DNA probes, such as pH> 9 solutions, caustic solutions (strong acids and bases, bleach, etc. )

Q3: While preparing standard solutions, in the pre-treatment steps why use the standard dilution buffer instead of DNase & RNase-free water to dilute?


Because the standard is relatively stable in the standard dilution buffer even after series dilution. But If diluted with water, the activity of the standard may change during the series dilution process. , resulting deviation of the standard.

Q4: How to adjust the appropriate microplate reader gain value?


Select auto-gain option if possible. If your microplate reader does not have auto-gain selection option, first set the middle value according to the gain value range; then adjust the gain value according to the fluorescence signal of the positive control (79 µL DNase & RNase-free water + 1 µL DNase I standard): reduce the gain value if the signal exceeds the upper limit of the instrument; Or increase the gain value appropriately if it is far below the upper limit of the instrument .

Q5: Why could a heavily contaminated sample lead to false negative results?


The criterion for judging RNase contamination is: RFU30 (sample to be tested) ≥ 2×RFU0 (sample to be tested), that is, the RFU value at 30 minutes of reaction is more than twice the RFU value at 0 minute. If the sample is seriously contaminated, it is possible that the reaction starts very quickly, so a very high RFU0 value is measured in a very short time, resulting in a false negative result of RFU30 (sample to be tested) < 2×RFU0 (sample to be tested) . At this time, the sample to be tested needs to be diluted with DNase & RNase-free water.

Q6: If the RFU value of the negative control is not 0, does it mean that the negative control is contaminated?


Uncertain. The negative control can also be detected with a low RFU signal (baseline value) but the value will not increase significantly along with the proceeding of the reaction. Generally, it is considered that the RFU30 is less than twice the RFU0 which indicates that the negative control is not contaminated.

Product details


Catalog Number Product Name Specification
HBP003002 RNase assay Kit (Fluorescence) 192 Tests
HBP003003 RNase assay Kit (Fluorescence) 48 Tests


References


[1] New England Biolabs DNase I User Guide.
[2] Kunitz M. Crystalline Desoxyribo - nuclease I. Isolation and General Pro - perties Spectrophotometric Method for the Measurement of Desoxyribonuclease Activity[J]. The Journal of General Physiology, 1950, 33 (4): 349-362.
[3] Thermo scientific RNase A (DNase and Protease-free) User Guide.
[4] Kunitz, M. A. A spectrophotometric method for the measurement of ribonuclease activity[J]. Journal of Biological Chemistry, 1946, 3 (2): 308-320.


Expected Usage


This kit is used to detect RNase contamination in samples.


Storage transportation conditions and expiry date 


1. Transported in -25 ~ -15℃.
2. The different components of the kit are stored separately according to the temperature:


Name Temperature
10×reaction solution -25~-15℃
RNA probe -25~-15℃
TE buffer -25~-15℃
RNase A standard (10mg/mL) -25~-15℃
Standard Dilution Buffer -25~-15℃
DNase&RNase-free water -25~30℃
DNase RNase away 2~30℃

3. Store the unopened kit at for 12 months.
4. Store the kit for 6 months after opening. It is recommended to aliquot the RNA probe solution according to the single use amount to avoid light and repeated freezing and thawing.


Also Testing for DNase Contamination?

Hzymes offers a companion DNase Assay Kit (Fluorescence) built on the same fluorescent-probe detection principle, with a limit of detection as low as 0.313 pg/mL (≈1.56×10⁻⁹ U/μL) for RNase A — also 1/8 that of the best competitor kit.

Message
Order Inquiry
Name *
Company *
Position *
Tel *
Mail *
Nation *

Do you have purchasing intention??

Requirement Description

Please contact on WhatsApp

Service Hotline: +86 400-808-5320

Large-scale production base: Building 6, Precision Medical Industry Base, Wuhan, China

Logistics & Supply Chain Center:417 Main St, Little Rock, AR 72201. United States.

Global Marketing Center: Hzymes Building, Fengxian District, Shanghai, China.

  • iso_copy_copy
  • iso_copy
  • iso
  • iso_copy_copy
  • iso_copy
  • iso
Contact Us

Service Hotline: +86 400-808-5320

Large-scale production base: Building 6, Precision Medical Industry Base, Wuhan, China.

Logistics & Supply Chain Center:417 Main St, Little Rock, AR 72201. United States.

Global Marketing Center: Hzymes Building, Fengxian District, Shanghai, China.

Copyright © Hzymes Biotechnology Co., Ltd. All Rights Reserved Web design

Site Map | Legal Notice | Privacy Policy |

Contact Us