HDL-Cholesterol Assay Kit (Direct Method)
HDL-Cholesterol Assay Kit (Direct Method)

  • 353
  • 354
80ml
HOEM0501-01
1000ml
HOEM0501-02
10L
HOEM0501-03

Cat. No: HOEM0501 

Storage and stability: Up to expiration date indicated on the label, when stored unopened at 2-8℃ and protected from light.  

Once opened, the reagents are stable for 28 days when refrigerated on the analyzer or refrigerator. Contamination of the reagents must be avoided.  

Do not freeze the reagents.  

Once dissolved, the calibrator are stable for 7 days at 2–8℃, the control are stable for 7 days at 2–8℃, do not freeze.

Technical Support

Features


1) Accuracy comparable to imported brands


2) Stable performance: 2–8℃ storage stable for 18 months, open-vial stability up to 30 days, 37℃ thermal stability up to 14 days


3) In-house production of key enzyme raw materials (CO, CE)


4) Flexible cooperation: from raw materials to reagents (OEM bulk packaging, formula cooperation)



Method Comparison


High accuracy, good correlation with imported brands



Good thermal stability, up to 14 days at 37℃



Good interference resistance, comparable to imported brands


Anti-Interference Ability


Interfering Substances

2-8℃ 14d

37'℃ 14d

Hzymes

Imported Brand

Hzymes

Imported Brand

Conjugated Bilirubin

-5. 80%

-7. 50%

-6. 80%

-7. 20%

Unconjugated Bilirubin

-5. 30%

2. 70%

-8. 50%

1. 50%

Hemoglobin

0. 00%

2. 60%

0. 00%

3. 00%

Ascorbic Acid

-8. 10%

-1. 20%

-8. 50%

-0. 30%

Lipemia

3. 60%

-2. 30%

-0. 30%

-5. 00%

Good interference resistance, comparable to imported brands



Application


In vitrote st for the quantitative determination of HDL-Cholesterol (HDL-C) concentration in serum on photometric systems. HDL cholesterol is inversely related to the risk of developing coronary artery disease. A low HDL/LDL cholesterol ratio is directly related to the risk of developing coronary artery disease. A high HDL cholesterol is associated with the "longevity" syndrome.



Principle


The System monitors the change in absorbance at 546nm. This change in absorbance is directly proportional to the concentration of cholesterol in the sample and is used by the System to calculate and express the HDL-cholesterol concentration.



Warnings and precautions


1. For in vitro diagnostic use.

 
2. Take the necessary precautions for the use of laboratory reagents.

 
3. Preservative contained. Do not swallow. Avoid contact with skin and mucous membranes.

 
4. Disposal of all waste material should be in accordance with local guidelines.

Message
Order Inquiry
Name *
Company *
Position *
Tel *
Mail *
Nation *

Do you have purchasing intention??

Requirement Description

Please contact on WhatsApp

Service Hotline: +86 400-808-5320

Large-scale production base: Building 6, Precision Medical Industry Base, Wuhan, China

Logistics & Supply Chain Center:417 Main St, Little Rock, AR 72201. United States.

Global Marketing Center: Hzymes Building, Fengxian District, Shanghai, China.

  • iso_copy_copy
  • iso_copy
  • iso
  • iso_copy_copy
  • iso_copy
  • iso
Contact Us

Service Hotline: +86 400-808-5320

Large-scale production base: Building 6, Precision Medical Industry Base, Wuhan, China.

Logistics & Supply Chain Center:417 Main St, Little Rock, AR 72201. United States.

Global Marketing Center: Hzymes Building, Fengxian District, Shanghai, China.

Copyright © Hzymes Biotechnology Co., Ltd. All Rights Reserved Web design

Site Map | Legal Notice | Privacy Policy |